PERFORMANCE AND TECHNICAL INFORMATION / 17
Measuring Temperature
Use a thermometer to measure the temperature of the
product water. It is most convenient to take this reading
when the product water flow rate is checked.
Checking System Performance
Procedure
The following procedure is summarized on the Performance
Worksheet printed on page 19. It details the measurement
and evaluation of the key aspects of Aqua-Cleer
®
/Good Water
Machine
system performance:
• Quality of water produced
• Quantity of water produced and stored
• Efficiency of operation
This procedure should be used to evaluate and record the
performance of a new system and to check the performance
of an operational system. The results of the new system
performance evaluation should be retained as a benchmark
of system performance in the years to come.
Checking Quality
Measure and record the TDS level of both the product water
and the feed water. Calculate the percent removal of TDS as
follows:
Removal = x 100%
As an example, consider a system which is producing 50
mg/L product water from a 1000 mg/L source:
[(1000 - 50) ÷ 1000] x 100% = 95% Removal
Checking Quantity Produced
Measure and record both the flow rate and temperature of the
product water. Record the temperature correction factor from
FIG. 12
Feed Water Correction Feed Water Correction Feed Water Correction
Temperature Factor Temperature Factor Temperature Factor
°F °C °F °C °F °C
36 2 0.33 52 11 0.56 68 20 0.83
38 3 0.34 54 12 0.59 70 21 0.87
40 4 0.37 56 13 0.63 72 22 0.90
42 6 0.40 58 14 0.65 74 23 0.94
44 7 0.43 60 16 0.69 76 24 0.96
46 8 0.46 62 17 0.72 77 25 1.00
48 9 0.50 64 18 0.76 78 26 1.03
50 10.0 0.52 66 19 0.79 80 27 1.06
TABLE 2 - TEMPERATURE CORRECTION FACTORS
(77°F (25°C) rating multiplied by correction factor equals capacity)
Rinse the cell cup twice with water to be tested, then
fill to the top.
Press the button on the front of the meter and read the
dial for the dissolved solids content of the product
water in parts per million (ppm).
Measuring Flow Rates
To measure flow rates, it is necessary to use a graduated
cylinder (100 ml suggested), and a watch or stopwatch with a
second hand. Measure and record the product and concen-
trate flow rates as follows:
Collect the water sample directly from the manifold
for exactly one minute or exactly two minutes.
Convert the measured flow rate to gallons per day (gpd)
as follows:
one minute sample: ml
two minute sample: ml
min.
2 min.
x 0.40 conversion = gpd
÷ 5 conversion = gpd
(Feed Water TDS - Product Water TDS)
(Feed Water TDS)